Optimized operation and calibration procedures for radical amplifier type dectors

dc.contributor.authorMihele, C.M.
dc.contributor.authorHastie, D.R.
dc.date.accessioned2010-05-25T19:49:14Z
dc.date.available2010-05-25T19:49:14Z
dc.date.issued2000
dc.description.abstractLaboratory experiments were conducted to improve the operation and calibration procedures for the ambient ROx radical detectors (ROx = HO2 + RO2 + OH + RO) based on chemical amplification. A new method for the experimental determination of the effective ratio between the absorption cross sections for water and molecular oxygen at 184.9 nm was developed and used to reduce the uncertainties associated with the peroxy radical source based on water photolysis. A novel peroxy radical source based on OCIO was developed and employed to confirm the ambient humidity effect on the chain length of the radical detector. The humidity effect on the chain length can be reduced by heating the walls of the reactor and decreasing the residence time of the air sample in the reactor to avoid the possible interferences from thermolabile species.en
dc.identifier.citationJ. Atm. Ocean Technol., 17, 788-794en
dc.identifier.urihttp://hdl.handle.net/10315/4158
dc.language.isoenen
dc.publisherAmerican Meterological Societyen
dc.rights.journalhttp://www.ametsoc.org/en
dc.titleOptimized operation and calibration procedures for radical amplifier type dectorsen
dc.typeArticleen

Files

Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
HAS054.pdf
Size:
449.62 KB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.83 KB
Format:
Item-specific license agreed upon to submission
Description: