mirage

Removal of phase artifacts from fMRI data using a Stockwell transform filter improves brain activity detection

DSpace/Manakin Repository

Removal of phase artifacts from fMRI data using a Stockwell transform filter improves brain activity detection

Show full item record

Title: Removal of phase artifacts from fMRI data using a Stockwell transform filter improves brain activity detection
Author: Goodyear, Bradley G.; Zhu, Hongmei; Brown, Robert A.; Mitchell, J. Ross
Abstract: A novel and automated technique is described for removing
fMRI image artifacts resulting from motion outside of the imaging
field of view. The technique is based on the Stockwell
transform (ST), a mathematical operation that provides the frequency
content at each time point within a time-varying signal.
Using this technique, 1D Fourier transforms (FTs) are performed
on raw image data to obtain phase profiles. The time
series of phase magnitude for each and every point in the phase
profile is then subjected to the ST to obtain a time-frequency
spectrum. The temporal location of an artifact is identified
based on the magnitude of a frequency component relative to
the median magnitude of that frequency’s occurrence over all
time points. After each artifact frequency is removed by replacing
its magnitude with the median magnitude, an inverse ST is
applied to regain the MR signal. Brain activity detection within
fMRI datasets is improved by significantly reducing image artifacts
that overlap anatomical regions of interest. The major
advantage of ST-filtering is that artifact frequencies may be
removed within a narrow time-window, while preserving the
frequency information at all other time points.
Subject: fMRI
artifacts
filtering
Stockwell transform
Fourier transform
motion
Type: Article
URI: http://hdl.handle.net/10315/928
Published: MAGNETIC RESONANCE IN MEDICINE
Series: 51
Citation: Goodyear, B.G., Zhu, H., Brown, R.A., Mitchell, J.R., “Removal of phase artifacts from fMRI data using a Stockwell transform filter improves brain activity detection”, Magn. Reson. Med., 51:16-21, 2004
ISSN: 0740-3194